Non-diagnostic ECG waveform input contract

The Python contract in openmed.multimodal.ecg.waveform normalizes caller-supplied ECG samples locally and deterministically. It performs no file reads, network calls, model inference, disease interpretation, or clinical decision. Its schema is openmed.multimodal.ecg.waveform.v1. This issue's implementation is the Python input contract; it adds no Swift provider or clinical workflow.

Every record, report, and validation error carries this boundary:

Non-diagnostic ECG input normalization only. No disease interpretation or clinical validation. Explicit reviewer confirmation is required before consequential use; never automatically trigger clinical decisions.

Canonical representation

  • EcgLead contains immutable tuples of millivolt samples and boolean validity masks (True means observed). Missing samples are exactly None with a false mask. An omitted conversion mask is derived from None. NaN, infinity, non-numbers, and conflicting masks are rejected; values are never interpolated.
  • Lead identifiers normalize whitespace and case to I, II, III, aVR, aVL, aVF, and V1 through V6. Records sort by this order and reject duplicate aliases. Unknown/custom leads must be mapped explicitly by the caller; no lead is inferred or derived from other leads.
  • EcgLead.from_samples requires explicit units: V, mV, uV, µV, μV, or adc. Voltage converts to millivolts. ADC conversion is (sample - baseline_counts) / gain_counts_per_mv; both calibration values must be supplied, finite, and gain must be positive. Physical voltage rejects ADC calibration to prevent double scaling. A source byte digest binds the caller's original header/calibration; the report does not copy that header.
  • EcgWaveform requires one to twelve unique aligned leads with equal sample counts and an explicit finite rate in (0, 100000] Hz. There are at most 1,000,000 samples across all leads. Each lead has at least one sample; a wholly missing lead is representable and is not a quality qualification.
  • Relative sample seconds are start_offset_seconds + index / sample_rate_hz. Start defaults to zero. Optional explicit offsets must match that regular grid within 1e-9 of a sample interval. Duration (count / rate) and final offset cannot exceed 86,400 seconds. Absolute timestamps, irregular timing, per-lead rates, gaps without masks, resampling, and asynchronous alignment are unsupported. Split or prepare these inputs explicitly before admission.
  • EcgProvenance accepts only a supported source format (array, dicom, edf, wfdb, hl7-aecg) and a lowercase source SHA-256 digest. Format declarations do not imply a reader, provider qualification, or clinical validation.

Synthetic example

import hashlib
from openmed.multimodal.ecg.waveform import EcgLead, EcgProvenance, EcgWaveform

record = EcgWaveform(
    leads=(EcgLead.from_samples(" ii ", (0, 1000, None), unit="uV"),),
    sample_rate_hz=250,
    provenance=EcgProvenance("array", hashlib.sha256(b"synthetic ECG v1").hexdigest()),
)
assert record.leads[0].samples_mv == (0.0, 1.0, None)
assert record.sample_offsets_seconds == (0.0, 0.004, 0.008)
assert record.to_report()["leads"] == [{"lead_id": "II", "missing_count": 1}]

Privacy and review boundary

Samples remain sensitive in memory. Normalization is not de-identification. repr, to_report, and deterministic to_json omit amplitudes, individual sample offsets, masks, source paths, header text, and patient metadata. Reports contain only schema/notice, review requirement, source format/digest, canonical units/leads, rate, sample count, and missing counts. Digest linkage can still be sensitive; callers must apply their access and retention policies. Do not log raw attributes or serialize the record with generic dataclass serializers.

Validation raises WaveformValidationError with static reason codes and actionable instructions, without reflecting supplied values or native exception context. Reasons cover unsupported leads/units/formats, invalid digest/rate, calibration/missingness/timing ambiguity, alignment, and resource limits.

record.require_reviewer_confirmation() fails closed. A caller must supply confirmed=True following explicit review before consequential use. The method does not approve or release an output, persist consent, or certify clinical safety. Downstream outputs must retain the notice and their own explicit review gate; the report always keeps review_required=True. There is no automatic clinical action or cloud fallback.

All tests use synthetic values and digest literals. No dataset, credentials, model assets, or additional dependencies are bundled.